8
SPECIFICATIONS
Measurement Range -40° to 1076°F (-40° to 580°C)
Distance-To-Spot (D:S) Ratio 12:1
Measurement Accuracy ±3°F (2°C) or ±2% of reading (whichever is greater)
for readings above 32°F (0°C); ±5°F (3°C) or ±2% of
reading (whichever is greater) for readings below
32°F (0°C)
Measurement Repeatability 1% of reading or 1°
Response Time 500 msec for 95% response
Emissivity Fixed at 0.95
Laser Class/Power/Wavelength
Class 3R/<3mW/665nm
Backlight Auto Dimming To half brightness after 20 seconds of front-panel
activity
Auto Power Off Trigger Inactivity for 60 seconds
Current Consumption <50mA, max
Operating Temperature 32° to 104°F (0° to 40°C)
@ <75% relative humidity (RH)
Storage Temperature -4° to 140°F (-20° to 60°C)
(without battery) @<85% RH
Power Source One “9V” battery (included)
Dimensions 6.7 x 5.3 x 1.4 in. (170 x 135 x 36mm)
Weight 5.6 oz. (158g)
MAINTENANCE AND
TROUBLESHOOTING TIPS
After subjecting the IRT to a sudden change in ambient temperature, you must
wait 30 minutes before measurements are within specifications.
The infrared lens on the front of the unit is the TS05’s most sensitive
component. To clean the lens, first use compressed air to blow off loose
particles and then brush any remaining debris away with a camel’s hair brush.
If necessary, carefully wipe the lens with a cotton swab moistened with water.