EasyManuals Logo
Home>Grant>Laboratory Equipment>SAP26

Grant SAP26 User Manual

Default Icon
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
JB Academy, JB Nova and SUB Aqua Pro 31394 V7
Detailed operating manual Page 18
www.grantinstruments.com
8 Calibration
8.1 Calibration options
Two calibration options exist:
Single point calibration (available on all models)
Dual point calibration (available on SUB Aqua Pro models only)
The quality of the calibration is highly dependent on:
Use of a suitable reference thermometer, ideally 10 times the accuracy
you are trying to achieve.
Performing a calibration in a stable ambient environment (+/-1ºC) free
from draughts or cooling air currents.
8.2 Single point calibration (and )
A single point calibration applies a single offset over the bath temperature curve. For this
reason the calibration temperature is usually the same as the intended working
temperature for the bath or particular experiment:

Table of Contents

Other manuals for Grant SAP26

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Grant SAP26 and is the answer not in the manual?

Grant SAP26 Specifications

General IconGeneral
BrandGrant
ModelSAP26
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals