EasyManuals Logo
Home>GW Instek>Test Equipment>GPT-700A Series

GW Instek GPT-700A Series User Manual

GW Instek GPT-700A Series
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
ELECTRICAL SAFETY TESTER
USER MANUAL
7
13) General
Power Source
AC100V, 120V, 220V, 230V±10% 50/60Hz
Operation
Environment
Indoor use, altitude up to 2000m.
Ambient Temperature 0 to 40.
Relative Humidity 80% (Maximum).
Installation category II
Pollution Degree 2
Storage temperature
& Humidity
-10 to 70.
70% (Maximum).
Accessories
GHT-205 × 1,
GTL-115 × 1(only for GPI-745A),
Instruction manual × 1,
Interface Manual × 1
Dimension 446(D) × 330(W) × 149(H) (m/m)
Weigh
Approx. 14.9 kgs
The main purpose provided by the series of instruments is for
Puncture Testing. The specification is not for continuous test. The
temperature of heat-sink is monitored. The test procedure will stop if
the heat-sink is too hot. Please refer to Figure 2-1.
Stop the instrument for 10 minutes when continuously proceeding
30~40mA withstanding test for one hour.
ELECTRICAL SAFETY TESTER
USER MANUAL
8
100 200 300 400 500 1000 2000 3000 4000 5000 6000 V
Figure 2-1
AC Hi-Pot
Spec 180sec
AC Hi-Pot Spec
Continuous
DC Hi-Pot Spec
Continuous
40
35
0
2
4
6
8
10
15
20
25
30
mA