Use a sample ID
The sample ID tag is used to associate measurements with a particular sample location. If assigned,
stored data will include the sample ID.
1. Push the key.
2. Select, create or delete a sample ID:
Option Description
Current ID Select an ID from a list. The current ID will be associated with sample data until a
different ID is selected.
Create a New Sample ID Enter a name for a new sample ID.
Delete Sample ID Delete an existing sample ID.
Use an operator ID
The operator ID tag associates measurements with an individual operator. All stored data will include
the operator ID.
1. Push the key.
2. Select, create or delete an operator ID:
Option Description
Current ID Select an ID from a list. The current ID will be associated with sample data until
a different ID is selected.
Create a New Operator ID Enter a name for a new operator ID (maximum 10 names can be entered).
Delete Operator ID Delete an existing operator ID.
Data management
About stored data
The following types of data are stored in the data log:
Sample measurements: stored automatically each time a sample is measured in the Press to
Read or Interval Mode. When the continuous measurement mode is used, data is stored only
when Store is selected.
Calibrations: stored only when Store is selected at the end of a calibration. Calibration data is also
stored in the IntelliCAL (R) probe.
Check standard measurements: stored automatically each time a check standard is measured (in
the Press to Read or Interval Mode).
When the data log becomes full (500 data points), the oldest data point is deleted when a new data
point is added. The entire data log can be deleted to remove data that has already been sent to a
printer or PC (
key > Delete Data Log). To prevent deletion of the data log by a user, use the
Security Options menu.
View stored data
The data log contains sample, calibration and check standard data. The most recent data point in the
data log is tagged as Data Point 001.
12
English

Table of Contents

Other manuals for Hach HQ411d

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach HQ411d and is the answer not in the manual?

Hach HQ411d Specifications

General IconGeneral
BrandHach
ModelHQ411d
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals