EasyManuals Logo
Home>Hach>Measuring Instruments>HQ440d

Hach HQ440d Basic User Manual

Hach HQ440d
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
1. Loosen the three battery cover screws and remove the battery cover (Figure 10).
Note: Do not remove the screws from the battery cover.
2. Remove the batteries.
3. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries. Make sure that the batteries
are installed in the correct polarity.
4. Replace the battery cover.
Note: Do not over-tighten the screws.
Figure 10 Battery replacement
Troubleshooting
Refer to the following table for common problem messages or symptoms, possible causes and
corrective actions.
Error/Warning Description Solution
Connect a Probe Probe disconnected or
connected improperly
Tighten the locking nut on the probe connector.
Disconnect the probe and then connect the probe
again
Software not updated to
most current version
To download the most current version of the
software, refer to the applicable product page on
the manufacturer's website.
Problem with probe Connect a different IntelliCAL probe to verify if
problem is with probe or meter
16 English

Table of Contents

Other manuals for Hach HQ440d

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach HQ440d and is the answer not in the manual?

Hach HQ440d Specifications

General IconGeneral
BrandHach
ModelHQ440d
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals