XNX Universal Transmitter Quick Start Guide
23
4.6.2 XNX EC Sensor Remote Mounting Kit
Theremotesensormountingkitisusedtoremotelymountthesensorfrom
thetransmitter.Toremotelymountthesensor,followtheprocedurebelow.
1. Unscrewtheweatherproofcover,loosentheretainerlockingscrew
andunscrewthesensorretainer.
2. Removethesensorbypullingwithouttwisting.
3. Plugtheremotesensorcableconnectorintothebottomofthe
transmitterandsecuretheretainer.
4. Routethecabletothelocationwheretheremotesensoristobe
mounted.
5. Optional:makealoopofcableatthejunctionbox.Thiswillprovide
someslackforanyfuturere-terminations.
6. Mounttheremotesensorjunctionbox.Allowenoughroombelowit
totthesensorandtheweatherproofcover.
7. Plugthesensorintothesocketatthebottomoftheterminalbox.
8. Fitthesensorretainer,tightenthelockingscrewandtthe
weatherproofcover.
9. Calibratethedetectorfollowingtheprocedure
sinSection8.1-
Calibration
.

Table of Contents

Other manuals for Honeywell XNX-AM Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX-AM Series and is the answer not in the manual?

Honeywell XNX-AM Series Specifications

General IconGeneral
BrandHoneywell
ModelXNX-AM Series
CategoryTransmitter
LanguageEnglish

Related product manuals