Function and Features
Copyright© Itech Electronics Co., Ltd. 22
Restored to Factory Setting (>Reset)
This option is used to restore all settings in the system menu to factory setting
values. Press [Enter] to restore to factory setting values. In this case, all set
values in the system will be restored to factory setting values, i.e., the (Def)
mark values.
Power On Parameter(>Power-on)
When the power-on parameter is set as Rst, at each time of power on, the set
parameters of the power supply will be 0V, 0A and the power rated value. The
setting parameter of load will be internal load rated value. The parameter
setting values under Setup and Function menus will also be restored to initial
values.
Rst will not initialize the system setting and configuration setting. If Sav0 is
selected, the parameters will be all setting values at the time of last power-off,
including output and input setting values of the power supply/load.
Trigger Mode (>Trigger)
Trigger is used for trigger the output of voltage, current and power, and there’re
three kinds of trigger options: Manual, Bus, and Ext. The default settings is
Manual
Manual: the trigger signal will be given by composite keys
[Shift]+[Enter](Trigger)
Bus: bus trigger mode.
Ext: external signal trigger.
Save Group Operation (Memory)
Power source can save 100sets parameters in nonvolatile memory by the save
group setting. This operation provides the customer with a convenient and quick
save/recall using condition.
GRP0: save(recall) power source parameters in 0-9 sets. Press
[Shift]+
[Recall]
(Save)+0-9 numeric keys([Recall] +0-9 numeric keys to recall the
parameters)
GRP1: save(recall) parameters in 10
th
-19
th
sets. Press [Shift]+
[Recall]
(Save)+numbers1-9 to save the parameters([Recall] +numbers1-9 to
recall parameters).Under this condition,number “1” represents to save or recall
the 10
th
parameters.Number “2” represents to save or recall the eleventh
parameter.GRP2-GRP9 can be understood in the same manner.
Key Sound Set (>Buzzer)
This item can set the buzzer state. On option indicates that when you push
buttons, the buzzer will sound. Off option indicates that the buzzer function is
disabled. Factory default is On option.
Communication Set (>Communication )
Under this item, you can set the concrete communication mode. This unit has
provided multiple communication interfaces: RS232/USB/GPIB/LAN/CAN.Of
which, the GPIB are addressable from 0-31. The baudrate options of RS232 are
4800, 9600, 19200, 38400, 57600, 115.2K. Data bit is 8bits. Parity bit has three
options: NONE, ODD, EVEN. Please ensure the configuration consistency
between our instrument and PC, so that you could have a successful