13
Quest syntax:ONPut:STATe?
Return parameter:0|1
[SOURce:]CURRent [:LEVel]
This order can set current value of power supply.
Order syntax:[SOURce:]CURRent [:LEVel] <NRf>
Parameter:MIN TO MAX|MIN|MAX
Unit:A mA
*RST value: MIN
Example: CURR 3A, CURR 30mA, CURR MAX, CURR MIN
Quest syntax:[SOURce:]CURRent [:LEVel]?
Parameter:[MIN|MAX]
Example:CURR?, CURR? MAX, CURR?MIN
Return parameter:<NR2>
[SOURce:]VOLTage[:LEVel]
This order can set voltage value of power supply.
Order syntax:[SOURce:]VOLTage[:LEVel] <NRf>
Parameter:MIN TO MAX|MIN|MAX
Unit:V mV kV
*RST value:MAX
Quest syntax:[SOURce:]VOLTage[:LEVel]?
Parameter:[MIN|MAX]
Return parameter:<NR2>
[SOURce:]VOLTage:PROTection[:LEVel]
This order can set voltage protection maximum level.
Order syntax:[SOURce:] VOLTage:PROTection[:LEVel] <NRf>
Parameter:MIN TO MAX|MIN|MAX
Unit:V mV
*RST value:
MAX
Example:VOLT:PROT 30V, VOLT PROT MAX
Quest syntax:[SOURce:] VOLTage:PROTection[:LEVel]?
Parameter:[MIN|MAX]
Example:VOLT:PROT?, VOLT PROT? MAX
Return parameter:<NR2>
4.4 Input measurement order
MEASure[:SCALar]:VOLTage[:DC]?
This order can get the input voltage of power supply.
Order syntax:MEASure[:SCALar]:VOLTage[:DC]?
Parameter:None
Return parameter:〈NR2〉
Return parameter unit:V
Example:MEAS:VOLT?