Automatic Test Function
Copyright © Itech Electronic Co., Ltd. 49
2. Set the maximum voltage value and press [Enter] to confirm.
MAX VOLT=120.00V
3. Set the maximum power value and press [Enter] to confirm.
MAX POWER=300.00W
4. Set the number of test steps, up to 10 steps, press [Enter] to confirm.
TEST STEP=10
5. Set the mode of the step, move the left and right cursors, when the mode
option flashes, it indicates that the mode is selected, press [Enter] to
confirm (such as select CC mode).
MODE CC CV CW CR
6. Set the current value and press [Enter] to confirm.
S01 CURR=**A
7. Set whether the step is short-circuited, move the left and right cursors to
select, and press [Enter] to confirm.
SHORT ON OFF
8. Set the type of value to be tested, move the left and right cursors to select,
and press [Enter] to confirm (such as selecting the value of the test as the
current value).
READBACK V A W
9. Set the upper limit current value and press [Enter] to confirm.
S01 MAX=**A
10. Set the lower limit current value and press [Enter] to confirm.
S01 MIN=**A
11. Set the delay time. That is, the test value is read after the input is set for a
certain period of time, and is used to test after the output is stable. The
delay range is 0.1~25 seconds.
S01 DELAY=**(S)
12. Repeat steps 5 through 11 to set the automatic test parameters for the next
steps.
13. Set the load voltage value. If no setting is required, set it to 0 and press
[Enter] to confirm.
START= **V
14. Set the test stop conditions. COMP means that the test stops after all tests
are completed, and FAILURE indicates that the test stops when an error
occurs.
STOP COMP FAILURE
15. Save the programmed file to EEPROM. You can save 50 groups of files. If
you want to put the edited file in the first group, press 1 and press [Enter]
to confirm.
SAVE PROGRAM =1 (1-50)
Note
In the process of editing the automatic test file, if an editing error
occurs, you can press the up arrow key to return to the previous step
to make changes without exiting the entire editing interface of the test
file.

Table of Contents

Other manuals for ITech IT8511+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8511+ and is the answer not in the manual?

ITech IT8511+ Specifications

General IconGeneral
BrandITech
ModelIT8511+
CategoryTest Equipment
LanguageEnglish

Related product manuals