Automatic Test Function
Copyright © Itech Electronic Co., Ltd. 48
SAVE 12
Note
The single-step parameter storage position is related to the number of
automatic test file storage group and the number of steps. If it is saved in
the first group, the single-step parameter storage position is consistent with
the number of steps. If it is saved in the second group, the single-step
parameter storage position is 1 merges the number of steps, for example,
the single-step 1, 2, and 3 are stored in 11, 12, and 13 respectively; if they
are stored in the third group, the single-step parameter storage position is 2
merges the number of steps, for example, the single-step 1, 2, and 3 are
saved in 21, 22, and 23, and so on.
Set the steps in CR/CW mode
Set the automatic test step in CW mode:
1. Press [CW], set the power value, and then press [Shift]+[CV] (Setup) to
enter the setting interface.
RANGE=150.00W
2. Set the maximum working power, and press [Enter] to confirm.
RANGE =20.00W
3. Set the upper limit of voltage, press [Enter] to confirm.
HIGH=120.00V
4. Set the lower limit of voltage, press [Enter] to confirm.
LOW=0.000V
5. The parameter settings are complete.
10.0000V 0.000A
0.00W W=1.00W
Set the automatic test step in CR mode:
1. Press [CR], set the resistance value, and then press [Shift]+[CV] (Setup)
to enter the setting interface.
RANGE=7500.0Ω
2. Set the maximum working resistance, and press [Enter] to confirm.
RANGE =2000Ω
3. Set the upper limit of voltage, press [Enter] to confirm.
HIGH=120.00V
4. Set the lower limit of voltage, press [Enter] to confirm.
LOW=0.000V
5. The parameter settings are complete.
10.0000V 0.000A
0.00W R=2.000Ω
4.2.2 Test Mode for IT8500
1. Press [Shift] + [6] (Prog) to set the maximum current value and press
[Enter] to confirm.
MAX CURR=30.00A