15
6
th
byte
The higher byte of current value.
7
th
byte
The highest byte of current value.
From 8
th
To 25
th
byte
System reserve
27
th
byte
Sum code
Represent current by 4 bytes of Hex. Lower bytes are in the front location, higher bytes are in
the later location. For example: current is 3.0000AHex code is 0X00007530NO. 4 bye is
0X30NO. 5 bye is 0X75NO. 6 bye is 0X00NO. 7 bye is 0X00
8. Set / Read voltage value of CV mode.2CH/2DH
1
st
byte
Start bit ( AAH )
2
nd
byte
Address (031,0XFF)
3
rd
byte
Command2CH/2DH
4
th
byte
The lowest byte of voltage value.
5
th
byte
The lower byte of voltage value.
6
th
byte
The higher byte of voltage value.
7
th
byte
The highest byte of voltage value.
From 8
th
to 25
th
byte
System reserve
26
th
byte
Sum code
Represent voltage by 4 bytes of Hex. Lower bytes are in the front location, higher bytes are in
the later location. For example :voltage is 16.000VHex code is 0X00003EB04th byte 0XB0
5TH byte is 0X3E6th byte is 0X007th bytes 0X00
9. Set / Read watt value of CW mode2EH/2FH
1
st
byte
Start bit ( AAH )
2
nd
byte
Address (031,0XFF)
3
rd
byte
Command2EH/2FH
4
th
byte
The lowest byte of max power value
5
th
byte
The lower byte of max power value
6
th
byte
The higher byte of max power value
7
th
byte
The highest byte of max power value
8
th
to 25
th
byte
System reserve
26
th
byte
Sum code
NOTE
NOTE
NOTE

Other manuals for ITech IT8512A+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8512A+ and is the answer not in the manual?

ITech IT8512A+ Specifications

General IconGeneral
BrandITech
ModelIT8512A+
CategoryTest Equipment
LanguageEnglish

Related product manuals