el 2110 5½ Digit Multimeter Calibration Manual Section 3:
2110-905-01 Rev. C / August 2013 3-5
Temperature accuracy verification
TEMPERATURE ACCURACY VERIFICATION
Range Test Point Accuracy Limits
Upper Limit Lower Limit
TC Type K 100.000 °C 101.000 °C 99.000 °C
Specifications valid after 2-hour warm-up at 5-1/2 digits:
a. ADC set for continuous trigger operation.
b. Input bias current < 30 pA at 25 °C.
c. Input protection 1000 V.
d. Measurement rate set to 10 PLC (fixed).
Capacitance accuracy verification
CAPACITANCE ACCURACY VERIFICATION
Range Test Point Accuracy Limits
Upper Limit Lower Limit
1 µF 1.00000 µF 1.01500 µF 0.98500 µF
µ
µ
µ
µ
µ
µ
µ
µ
Specifications valid after 2-hour warm-up at 5-1/2 digits:
a. ADC set for continuous trigger operation.
b. Input test current: 10 uA (1 nF to 10 nF), 100 µA (100 nF to 10 µF),
and 1 mA (10 µF to 100 µF).
c. Input protection 1000 V all ranges and 1 mA (10
µ