REMOTE CONTROL
59
Parameter/Return
<NRf>
0~105% of the rated output voltage in volts.
MIN
Minimum voltage level
MAX
Maximum voltage level
Example SOUR:VOLT:LEV:IMM:AMPL 10
Sets the voltage level to 10 volts.
[SOURce:]VOLTage[:LEVel]:TRIGgered
[:AMPLitude]
Description
Sets or queries the voltage level in volts when a
software trigger has been generated.
Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
[MIN|MAX]
Parameter/Return
<NRf>
0%~105% of the rated voltage output in volts.
MIN
Minimum current level.
MAX
Maximum current level.
Example SOUR:VOLT:LEV:TRIG:AMPL 10
Sets the voltage level to 10 volts when a software
trigger is generated.
[SOURce:]VOLTage:PROTection[:LEVel]
Description
Sets or queries the overvoltage protection level.
Syntax [SOURce:]VOLTage:PROTection[:LEVel]
{<NRf>|MIN|MAX}
Query Syntax [SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Parameter/Return
<NRf>
10%~110% of the rated output voltage in
volts.
MIN
Minimum OVP level
MAX
Maximum OVP level
Example SOUR:VOLT:PROT:LEV MAX
Sets the OVP level to its maximum.
Set
Query
Set
Query