Series 3700A Firmware Release Notes Version 1.53 Release - Non-critical fixes
3706A-850SV1.55a / June 2017
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
Unexpected behavior with scan.background() and schedule.alarm commands.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When run using Java plug-in version 7u25 or later, Series 3700A web applications will eventually stop
working after it is re-loaded.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When running a Series 3700A web application, a JAVA security prompt may ask for your confirmation
for running the application. An incorrect application name was shown on the security dialog window.
Workaround:
None.
Resolution:
This issue has been corrected.