Series 3700A Firmware Release Notes Version 1.51 Release - Enhancements
3706A-850SV1.55a / June 2017
17
Symptom:
It was observed that reading large amounts of data using VXI would occasionally cause the data to be
terminated early.
Resolution:
During heavy amounts of reading the VXI driver would set the EOI bit too early terminating the data
transfer. This was corrected.
Enhancements
None.
Noncritical fixes
Models affected:
3706A, 3706A-S, 3706A-NFP, 3706A-SNFP
Symptom:
The login window of the Java applet was not sized properly.
Resolution:
The resolution of the login window was set to correct values.
Models affected:
3706A, 3706A-S, 3706A-NFP, 3706A-SNFP
Symptom:
channel.calibration.verifydate returns a misleading error for cards that do not support calibration.
Resolution:
For cards that do not support calibration the error was changed to “Card not supported”.
PR45677