Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
If the user edits a script in TSB Embedded and clicks a non-applet link such as Sync Config prior to
saving, when they return to TSB Embedded the changes are gone and no script is displayed in the
editor.
Workaround:
Be sure to save your work before selecting another link.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
Upgrading or Downgrading to firmware version 1.54a might cause the instrument to lock-up.
Workaround:
Power cycle the instrument. On very rare occasions the instrument has been permanently disabled
and needs to be serviced.