8 Service, maintenance, disposal
8.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild
soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue
sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods must be removed immediately.
8.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN.
Before opening, disconnect from power supply.
8.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law
of the location where the appliance is used.
9 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The
weighing process must then be restarted from the beginning.
Help:
Fault
Possible cause
The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
The displayed weight is permanently
changing
Draught/air movement
-
Table/floor vibrations
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
The weighing value is obviously
The display of the balance is not at zero
wrong
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging (choose different
location/switch off interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error message remains inform
manufacturer.