1 Introduction
1.1 Features
Full Tare
Memory function with indicator
Stable indicator
Negative value indication
Two types of Digital Auto Adjusting
Solder Pads to prevent end-user internal adjusting
Low battery indicator
Auto off function (battery mode only)
AC adaptable
Display segment test function
Overload indication / protection
1.2 Digital auto adjusting procedure
1. Turn balance on and allow the unit to acclimatise and stabilise for two minutes.
2. Press and hold the MODE key until the display shows a flashing weight reading.
The flashing weight reading indicates the correct weight that must be placed on the
platform.
3. With the display still flashing, press the OFF key. The display will now show CAL 0.
Then gently place the correct adjusting weight on the centre of the platform.
4. The display will now show CAL F and then return to normal operating mode. The
adjusting is complete when the display correctly shows the weight that is placed on
the platform and the display has stabilised.
5. Remove the adjusting weight and press the TARE key to reset the zero point..
In case CAL E is display instead of CAL F, this indicates an adjusting procedure error.
Turn the balance off and then on and repeat the procedure.