EasyManuals Logo
Home>KERN>Scales>EW - 600 - 2M

KERN EW - 600 - 2M User Manual

KERN EW - 600 - 2M
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
EW/EG-KB-e-9811 15
4.3 How to adjust overload stopper
1. Apply the pan base on the mechanism.
Then slide out stoppers by loosening two fixing screws with a hexagon wrench of 3 x 3.
2. Press the ON/OFF key to on while pressing F key and T key together
Release both keys when DJr____ appears.
3. Apply reference weights up to the Stopper Setting Value specified in the
table in page 15*1.
Press the T key to clear the display to zero (0).
4. Insert stoppers until they slightly touch corresponding guide link.
Holding the stopper with a screw driver
or the like, fix the stopper by driving
the cap screw securely.
5. The adjustment is good if display shows
between -5g and -50g.
Even the display shows "0", if the guide
link touches (stopped by) stoppers by increasing reference weights up to
Maximum specified in the table 15*2, the adjustments is good.
HOW TO KNOW WHEN LINK TOUCHES STOPPER
While increasing weights, when the guide link touches the stopper, the
display will not show as much value as increased weight.
This indicates the guide link has touched the stopper.
6. How to correct the setting
If the display value in above 5. Is less than -50g, such as -100g for
example, slide out stoppers a little from the guide link.
If the display value in above 5. Is over -5g, such as +5g for example,
insert the stoppers slightly more into the guide link.