EW/EG-SU-e-9811 20
1. How to remove weighing mechanism
(1) Remove the upper case referring to page 9.
(2) Remove three mechanism fixing nuts by a nut driver of opp. 7mm.
(3) Remove wires to the tuning-fork sensor, by welding solders at DJD circuit board.
(4) Remove the mechanism unit by holding at the lower case