FTC-BA-e-0720
31
Settings in the menu
User manual ME-Nr.22017145 01/07
4.3.5 RESTART – automatic saving of zero point and tare value
4.3.6 FILTER – adaptation to the ambient conditions and the weighing type
4.3.7 RESET – resetting scale settings to factory settings
ON/OFF When the Restart function is activated, the last zero point and tare value are saved.
After switching off / on or after a power interruption, the device continues to work with
the saved zero point and tare value.
VIbrAt Adaptation to the ambient conditions
LOW
• Very steady and stable environment. The scale works very quickly, but is very
sensitive to external influences.
MEd
• Normal environment. The scale operates at medium speed.
HIGH
• Restless environment. The scale works more slowly, but is insensitive to external
influences.
PrOCESS Adaptation to the weighing process
UNIVEr
• Universal setting for all weighing samples and normal weighing goods
dOSING
• Dispensing liquid or powdery weighing samples
StAbILI Adjusting the weighing speed
FASt
• The scale operates very fast.
StAndrd
• The scale operates at medium speed.
PrECISE
• The scale operates with the greatest possible reproducibility.
The slower the scale works, the greater the reproducibility of the weighing results.
SUrE? Confirmation inquiry
• Reset the scale settings to factory settings with
• Do not reset scale settings with