EasyManuals Logo
Home>KERN>Scales>FTC

KERN FTC User Manual

KERN FTC
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #32 background imageLoading...
Page #32 background image
Settings in the menuFTC-BA-e-0720
32User manual ME-Nr.22017145 01/07
4.4 Application settings (APPLICATION)
4.4.1 COUNT – settings for counting
VAr-SPL Adaptation of the reference quantity
ON
The reference quantity can be changed in operating mode
OFF
Counting only with defined reference quantities
SPL-qtY Reference quantity
Sq1 ... Sq5
Define 5 fixed reference quantities
Min.reFW Monitoring the minimum reference weight
OFF
No monitoring of the minimum reference weight
97.5, 99.0,
99.5
Monitoring the minimum reference weight so that a counting accuracy of
97.5 %, 99.0 % or 99.5 % is achieved
rEF.OPt Optimizing the average piece weight
OFF
No reference optimization
AUtO
Automatic reference optimization

Other manuals for KERN FTC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN FTC and is the answer not in the manual?

KERN FTC Specifications

General IconGeneral
BrandKERN
ModelFTC
CategoryScales
LanguageEnglish

Related product manuals