19
Product Service 1-800-522-7658, International 816-333-8811
Chapter 3: Getting Started
Vent Installation – Acid and Standard
Base Cabinet
To ventilate the Acid Storage Cabinet and the Standard
Base Storage Cabinet properly, first decide which of the
two vent connections in the back of the cabinet that you
intend to use. The lower vent connection should be
used for heavier than air vapors and the upper vent
connection should be used for lighter than air vapors to
obtain the best results within the cabinet structure.
Vent Kit, P/N 35911, allows the Acid Storage Cabinet
or Standard Base Cabinet to be vented directly into a
Protector® Laboratory Hood. Run the vent line
through the work surface into the fume hood or connect
through separate duct system to the outside. The vent is
sized for 2" diameter nominal plastic pipe.