7 Single-point calibration, C300
CAL1: With CAL1 (single-point calibration) you set the offset
(error correction + height correction, see table) for the air
pressure. The offset is shown in the lower part of the display.
Maximum offset is +/- 250.0 mbar / hPa, or 10.00 inHg, or 5.00
psi, and be entered in steps of 0.1 mbar / hPa, or 0.01 bar, or
0.01 inHg, or 0.01 psi. In this context the signs, and the figures
before and after the decimal point are set individually, starting
with the sign. The offset is active across the whole measuring
range.
The original works settings (elevation head = 0m) can be
achieved by setting the offset to 0.0.
Two-point calibration is not possible with the C300.