EasyManuals Logo
Home>Metrix>Test Equipment>PX 110

Metrix PX 110 User Manual

Metrix PX 110
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
33Wattmeters
4.3 Environmental parameters
Climatic conditions
Temperature in °C
95
90
75
45
10
-40
0
20
26
35
50
70
Relative humidity in % RH
Note: The operating range (2) is reduced to 40°C with the mains adapter accessory.
Altitude
- Operation: 2,000 m
- Storage 10,000 m
Impermeability:protection index IP 54 (according to EN 60529)
(IP2X electrical protection for the terminals)
4.4 Compliance with norms
Electrical safety
(as per EN 61010-1)
- Dual insulation:
- Installation category: III
- Pollution level: 2
- Rated voltage: 600 V
Electromagnetic compatibility (EN 61326-1)
Electrostatic discharge (IEC 61000-4-2)
- 4 kV contact, 8 kV air
- aptitude criterion A, industrial site
Electromagnetic field (IEC 61000-4-3)
- 10 V/m
- aptitude criterion A, industrial site
Bursts (IEC 61000-4-4)
- 2 kV Level III
- aptitude criterion B, industrial site
Shock wave (IEC 61000-4-5)
- 1 kV differential mode, 2 kV common mode
- aptitude criterion A, industrial site
Conducted interference (IEC 61000-4-6)
-3 V
- aptitude criterion A, industrial site
Mechanical resistance
- Operating position: any - Impacts: 100 g (test as per IEC 68-2-27)
- Free fall: 1 m (test as per IEC 68-2-32) - Vibrations: 0.15 mm (test as per IEC 68-2-6)
Auto-extinguishability
- Housing: V0
- Display window: V0
Chapter IV
Reference range
Operating range
Storage range (without battery)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775

Other manuals for Metrix PX 110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrix PX 110 and is the answer not in the manual?

Metrix PX 110 Specifications

General IconGeneral
BrandMetrix
ModelPX 110
CategoryTest Equipment
LanguageEnglish

Related product manuals