EasyManuals Logo
Home>Mindray>Medical Equipment>BeneHeart C2

Mindray BeneHeart C2 User Manual

Mindray BeneHeart C2
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
A - 7
A.11 ECG Specifications (for Equipment Configured with the Screen)
A.12 Electrode Pads Specifications
ECG inputs Multifunction electrode pads
Gain Auto
Sweep speed 25 mm/s, error no more than ± 5%
Common mode rejection >90 dB
Recovery time <2.5 s (after defibrillation)
Electrode Pads MR60 MR61 MR63 MR62
Electrode shape Oval
Cable length 1.2 m preconnectable
Total area 148 cm
2
75 cm
2
75 cm
2
126 cm
2
Adhesive Area 145 cm
2
74 cm
2
74 cm
2
123 cm
2
Maximum Number of
Defibrillation Shocks
Up to 50 shocks (360J monophasic and biphasic)
Shelf life (with sealed package) 36 months 60 months
Storage Condition 0ºC to 50ºC 15ºC to 35ºC
The shelf life assumes a storage
temperature of 25ºC. Storage
temperature above 25ºC will reduce
the shelf life.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart C2 and is the answer not in the manual?

Mindray BeneHeart C2 Specifications

General IconGeneral
BrandMindray
ModelBeneHeart C2
CategoryMedical Equipment
LanguageEnglish

Related product manuals