Safety Precautions 1-3
1.4 Safety Precautions
Please observe the following precautions to ensure patient and operator’s safety when using this
system.
DANGER:
Do not operate this system and probes in an atmosphere containing
flammable gasses or liquids such as anesthetic gasses, hydrogen,
and ethanol, because there is danger of explosion.
WARNING:
1.
Do connect the power plug of this system to wall receptacles that
meet the ratings indicated on the rating nameplate. If adapters or
multifunctional receptacles are used, it may cause the leakage
current to exceed the safety requirement.
2.
In the environment that patient is 1.5 meters around, connect
peripherals to the auxiliary power outlet which is capable of
isolation protection, or power the peripherals by auxiliary output
cable or isolation transformer complied with IEC 60601-1-2:2014 or
the power input of the same safety level.
3.
DO NOT use power supply of different phases to power
peripherals, like power supply of air-conditioning.
4.
When using peripherals not powered by the auxiliary output of
the ultrasound system, or using peripherals other than permitted
by Mindray, make sure the overall leakage current of peripherals
and the ultrasound system meets the requirement of the local
medical device electrical regulation (like enclosure leakage
current should be no more than 500uA of IEC 60601-1-2:2014),
and the responsibility is held by the user.
5.
Connect the grounding conductor before turning ON the system.
Disconnect the grounding cable after turning OFF the system.
Otherwise, electric shock may result.
6.
For the connection of power and grounding, follow the appropriate
procedures described in this operator’s manual. Otherwise, there is
risk of electric shock. Do not connect the grounding cable to a gas
pipe or water pipe; otherwise, improper grounding may result or a
gas explosion may occur.
7.
Before cleaning the system, disconnect the power cord from the
outlet. Otherwise, system failure and electric shock may result.
8.
This system is not water-proof. Do not use this system in any place
where water leakage may occur. If any water is sprayed on or into
the system, electric shock may result. If water is accidentally
sprayed on or into the system, contact Mindray Customer Service
Department or sales representative.
9.
DO NOT use a probe that has a damaged, scratched surface, or
exposed wiring of any kind. Immediately stop using the probe and
contact Mindray Customer Service Department or sales
representative. There is risk of electric shock if using a damaged or
scratched probe.