14
ULTRASONIC POWER MEASUREMENT
POWER RANGE
0 to 30 W
RESOLUTION
Model
Grams
Watts
UPM-DT-10PA
1 mg
20 mW
UPM-DT-1PA
0.1 mg
2 mW
UPM-DT-1000PA
0.1 mg
1 mW
ACCURACY
±3% + One Count Reading
DUT OPERATING
FREQUENCY
0.5 to 10 MHz
DUT MAXIMUM
TRANSDUCER SIZE
4.50 Inches (114.3 mm)
TEST MEDIUM
Degassed Water @ 24.0 ±3.0 °C (75.2 ±5.4 °F)
ELECTRICAL
POWER SUPPLY
12 VDC, 500 mA
PHYSICAL & ENVIRONMENTAL
DISPLAY
7 Segment, Backlit LCD
CONSTRUCTION
TANK LINING
0.5 Inch (12.7
mm) thick
Neoprene
CARRYING CASE
Aluminum
SIZE
OVERALL (Including case):
11 x 15 x 10 Inches
(280 x 381 x 254 mm)
WEIGHT
OVERALL (Including case):
< 16.0 Lbs (7.25 kg)
OPERATING RANGE
15 to 30 °C
(59 to 86 °F)
STORAGE RANGE
-40 to 60 °C
(-40 to 140 °F)
SPECIFICATIONS