22
BX53/51/41-P
Remove specimen, test plate, compensator, etc. from the light
path.
1. Swing out the top lens of the condenser, and engage the 10X objective.
2. Insert the rotatable analyzer into the light path and set vibration direction
scale at the 0° position. Clamp with the clamping knob. (Fig. 30)
3. Set the polarizer scale at the 0° position.
4. Loosen the polarizer clamping knob @. Rotate the polarizer rotation
dial ², until complete extinction is obtained. At this point, tighten the
clamping knob @. (Fig. 30)
Fig. 30
2
Adjusting for Extinction (Fig. 30)
@²
CAUTION

Questions and Answers:

Olympus BX53-P Specifications

General IconGeneral
FocusingCoaxial coarse and fine focusing knobs
Optical SystemUIS2 optical system
IlluminationLED or Halogen
Observation MethodBrightfield, Darkfield, Phase Contrast, Fluorescence
StageMechanical stage with XY movement
Camera CompatibilityCompatible with Olympus digital cameras
NosepieceRevolving nosepiece
CondenserAbbe condenser
Power SupplyAC 100-240V

Summary

Nomenclature

Assembly and Setup

Assembly Diagram

Diagram illustrating the order and components for microscope assembly.

Detailed Assembly Procedure

Step-by-step guide for assembling microscope components like stage, condenser, and nosepiece.

Microscope Controls Overview

Using the Controls

Stage Operation

Instructions for specimen placement on the stage.

Stage Rotation

Procedure for rotating the stage horizontally and using click-stop features.

Stage Height Adjustment

How to adjust the stage height for accommodating taller specimens.

Intermediate Polarizing Attachment

Guide to using the Bertrand lens and analyzer for polarized light observation.

Polarized Light Observation

Adjustments Before Observation

Essential optical axis adjustments, including polarizer and analyzer setup for extinction.

Orthoscopic Observation

Procedure for performing orthoscopic observation using polarized light.

Conoscopic Observation

Steps for achieving conoscopic observation and focusing the conoscopic image.

Technical Specifications

Optical Characteristics

Troubleshooting Guide

Optical System Issues

Diagnosing and resolving common issues related to the microscope's optical system.