EasyManuals Logo
Home>Olympus>Medical Equipment>OSF-3

Olympus OSF-3 User Manual

Olympus OSF-3
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background image
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
35
OSF-3 ENDOSCOPE REPROCESSING MANUAL
2. Carefully coil the endoscope’s Insertion Tube, and Universal Cord, and
completely immerse the endoscope in the detergent solution.
Do not coil Insertion Tube and Universal Cord with a
diameter of less than 12 cm. Equipment damage can result.
3. Soak the endoscope for 10 hours at the temperature recommended by the
detergent manufacturer.
4. Remove the endoscope from detergent solution.
After soaking the endoscope, manually clean the endoscope following the
standard procedure described in Section 3.5, “Manual Cleaning”, then disinfect
or sterilize the endoscope following the procedures described in Section 3.6,
“High Level Disinfection” or Section 3.8, “Sterilization”.
3.6 High Level Disinfection
All disinfection steps should be performed with the
endoscope and all equipment completely immersed.
Otherwise, disinfectant solution may not adequately contact
all surfaces of the equipment. As a result, the effectiveness
of disinfection may be reduced.
After manual cleaning, disinfect the endoscope according to the procedure
described below.
Equipment Needed
Prepare the following equipment:
Personal Protective Equipment
Lint-free Cloth
Large Basin
Disinfectant Solution
All-channel Irrigator (CW-3)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus OSF-3 and is the answer not in the manual?

Olympus OSF-3 Specifications

General IconGeneral
BrandOlympus
ModelOSF-3
CategoryMedical Equipment
LanguageEnglish

Related product manuals