SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range:
J-TYPE -200°C to 1050°C, (-328°F to 1922°F)
K-TYPE -200°C to 1370°C, (-328°F to 2498°F)
Resolution: 0.1°C or 0.2°F
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
±(0.05% rdg + 0.3°C) -50°C to 1370°C
±(0.05% rdg + 0.7°C) -50°C to -200°C
±(0.05% rdg + 0.6°F) -58°F to 2498°F
±(0.05% rdg + 1.4°F) -58°F to -328°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Maximum Differential Common Mode Voltage (Maximum Voltage be-
tween T1 and T2 during measurement): 1volt.
Reading Rate: one time per second.
Input Connector: Accepts SMP male miniature thermocouple connectors.