15
EN
9. Technical Data
Note: Subject to technical modification without prior notice.
Please report to the manufacturer and the competent authority of the Member State in which you are
established about any serious incident that has occurred in relation to this device.
Product Category Body Composition Analyzers
Product Description Body Composition Monitor
Model (Code) BF511 (HBF-511T-E/HBF-511B-E)
Display
Body Weight:
0 to 150 kg with an increment of 0.1 kg
(0.0 to 330.0 lb with an increment of 0.2 lb)
Body Fat percentage:
5.0 to 60.0% with an increment of 0.1%
Skeletal Muscle percentage:
5.0 to 50.0% with an increment of 0.1%
BMI:
7.0 to 90.0 with an increment of 0.1
Resting Metabolism:
385 to 3999 kcal with an increment of 1 kcal
Visceral Fat Level:
30 levels with an increment of 1 level
BMI classification:
– (Underweight) / 0 (Normal) / + (Overweight) / ++ (Obese) 4 levels with 3 sublevels each
Body fat percentage and Skeletal muscle percentage classification:
– (Low) / 0 (Normal) / + (High) / ++ (Very High) 4 levels with 3 sublevels each
Visceral fat level classification:
0 (Normal) / + (High) / ++ (Very High) 3 levels with 3 sublevels each
Setting Items Height: 100.0 to 199.5 cm (3' 4" to 6' 6 3/4")
Age: 6 to 80 years old
Gender: Male/Female
* Measurement Units: kg (cm) / lb (feet•inch)
* The age range for the Body Fat percentage, Body Fat percentage classification,
Skeletal Muscle percentage, BMI and BMI classification and Resting
Metabolism rate is 6 to 80 years old.
* The age range for Visceral Fat level, Visceral Fat level classification and
Skeletal Muscle percentage classification is 18 to 80 years old.
Weight Accuracy
0.0 kg to 40.0 kg:
±
0.4 kg
(0.0 lb to 88.2 lb:
±
0.88 lb)
40.0 kg to 150.0 kg:
±
1%
(88.2 lb to 330.0 lb:
±
1%)
Accuracy (S.E.E.)
Body Fat percentage:
3.5%
Skeletal Muscle percentage:
3.5%
Visceral Fat Level:
3 levels
Durable Period
5 years
Power Supply
4 AA batteries (R6) (You may also use AA alkaline batteries (LR6).)
Battery Life
Approximately 1 year (when manganese batteries are used with four measurements
a day)
Operating Temperature /
Humidity / Air Pressure
+10 to 40°C / 30 to 85% RH (non-condensing) / 700 to 1060 hPa
Storage and Transport
Temperature / Humidity /
Air Pressure
-20 to 60°C / 10 to 95% RH (non-condensing) / 700 to 1060 hPa
Protection Against
Electric Shock
Internally powered ME equipment
Applied Part
Type BF (Grip Electrodes, Foot Electrodes)
IP Classification
IP21
*IP classification is degrees of protection provided by IEC 60529.
This unit is protected against solid foreign objects of diameter 12.5mm such as a finger and greater. This unit is
protected against the ingress of vertically falling water drops.
Weight
Approximately 2.2 kg (4.85 lb) (including batteries)
External Dimensions
Display unit:
Approx. 300 (W) × 35 (H) × 147 mm (D)
(Approx. 11 3/4" (W) x 1 3/8" (H) x 5 3/4" (D))
Main unit:
Approx. 303 (W) × 55 (H) × 327 mm (D)
(Approx. 11 7/8" (W) x 2 1/8" (H) x 12 7/8" (D))
Package Contents
Body Composition Monitor, 4 AA manganese batteries (R6), instruction manual,
Guarantee Card