45
10. Specifications
Model BP7450 HEM-7342T-Z
Display LCD digital display
Cuff pressure range 0 to 299 mmHg
Blood pressure
measurement range
SYS: 60 to 260 mmHg
DIA: 40 to 215 mmHg
Pulse measurement
range
40 to 180 beats / min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation Automatic by electric pump
Deflation Automatic pressure release valve
Measurement method Oscillometric method
Transmission method
Bluetooth
®
low energy technology
Wireless
communication
Frequency range: 2.4 GHz
(2400 - 2483.5 MHz)
Modulation: GFSK
Effective radiated power: < 20 dBm
Operation mode Continuous operation
IP classification
Monitor: IP 20
AC adapter: IP21
Rating DC6 V 4.0 W
Power source
4 “AA” batteries 1.5 V or AC adapter (INPUT
AC100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Battery life
Approximately 700 measurements (using
new alkaline batteries)
Durable period
(Service life)
Monitor: 5 years
Arm cuff: 5 years
AC adapter: 5 years
Operating conditions
+50°F to +104°F (+10°C to +40°C) / 15 to
90% RH (non-condensing) / 800 to 1060 hPa
Storage / Transport
conditions
-4°F to +140°F (-20°C to +60°C) / 10 to 90%
RH (non-condensing)