Troubleshooting
In case of any of the below problems occur during use, first check that
no other electrical device is within 30 cm. If the problem persists, please
refer to the table below.
The device does not switch on
Check that the power plug is properly fitted to the power outlet and the
power cord is connected to the socket of the compressor.
Make sure the power button is in on position.
The device switches on but does not nebulize
Make sure the vaporizer head is attached to the inhalation top of the
nebulizer kit.
Make sure the air tube is not squashed or crooked.
Check the air filter for blockage and dirt. Replace if necessary.
Check that sufficient amount of medication has been put into the
nebulizer kit.
The device suddenly stops working during operation.
The thermal cut-out has shut the device down for one of the following
reasons:
- the device was working in an environment with temperatures higher
than 40°C;
- the ventilation slots were covered.
Do not attempt to repair the device. Do not open and/or tamper with the
device. No parts of the device are user serviceable. Return the device to
an authorized OMRON retail outlet or distributor.
Please report to the manufacturer and the competent authority of the
Member State in which you are established about any serious incident
that has occurred in relation to this device.