11. GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON battery-operated BPM is intended for use in the electromagnetic environment specified
below. The customer or the user of this OMRON battery-operated BPM should assure that it is used in
such environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact discharge:
±6 kV
Air discharge:
±8 kV
Contact discharge:
±8 kV
Air discharge:
±15 kV
The relative humidity should be
at least 5 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
Not Applicable. Not Applicable.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not Applicable. Not Applicable.
Voltage dips, short
interruptions and
voltage variations
on power supply
inputlines
IEC 61000-4-11
<5 % U
T (>95 % dip in
U
T)
for 0.5 cycle
Not Applicable.
Not Applicable.
40 % U
T (60 % dip in
U
T)
for 5 cycles
Not Applicable.
70 % U
T (30 % dip in
U
T)
for 25 cycles
Not Applicable.
<5 % U
T (>95 % dip in
U
T)
for 5 sec.
Not Applicable.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 30 A/m
Magnetic fields from common
appliances are not expected to
affect the device.
Note: U
T is the A.C. mains voltage prior to application of the test level.