Technical data for the OMRON NE-C801 Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:
Result of cascade impactor measurements for particle size**
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering,
Doshisha University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009
Note: Performance may vary with drugs such as suspensions or high viscosity.
See drug supplier’s data sheet for further details.
Particle Size: **MMAD Approx. 3µm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank
Capacity:
7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Nebulization Rate: *Approx. 0.3ml/min (by weight loss)
Aerosol Output: **0.47ml (2ml, 1%NaF)
Aerosol Output Rate: **0.06ml/min (2ml, 1%NaF)