How to Select and Perform a Test
This is the basic procedure for all tests. For specific test details
please refer to the Test Instructions.
Lumiso
Pooltest
A Palintest Product
Chl(0–10)
Chl(0–5)
pH
12:28:49
18-01-2021
Lumiso
Pooltest
A Palintest Product
12:28:49
18-01-2021
Chl(0–10)
Chl(0–5)
pH
The test selection menu
appears when first starting
up Lumiso.
From most other screens,
pressing “back” once or
twice, will take you back to
the test selection menu.
Select a test, using the up and
down keys until the desired
test is highlighted.
These tube icons
will usually appear.
These indicate that the left
key is for “blanking” and
the right key is for “reading”
Please Note: The read icon
only appears after the blanking stage
has been completed. Lumiso will require
blanking to be repeated after 50
minutes or if it registers a significant
change in environmental temperature.
1
2
Blank Read

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the palintest Lumiso Pooltest 3 and is the answer not in the manual?

palintest Lumiso Pooltest 3 Specifications

General IconGeneral
Brandpalintest
ModelLumiso Pooltest 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals