24
The instrument is proof against:
- temperature of an ambient air from
minus 20
0
С up to plus 60
0
С with
indication of measurement results on
LCD;
- temperature of an ambient air from
minus 40 up to minus 20°С without
indication of measurement results on
LCD but with recording of measurement
results in a non-volatile memory;
- relative humidity of an ambient air up
to 98 % at 35
0
С;
- atmosphere pressure from 84 up to
106,7 kPa.
25
The instrument is proof against:
- sinusoidal vibration in the frequency
range 10 – 55 Hz and bias amplitude for
frequencies lower than the transition
frequency 0.35 mm;
- shocks with acceleration 100 m/s
2
,
duration of shock pulse 2-50 ms, shock
rate is 60 - 180 shocks per minute.
26
The instrument meets drop test against
a concrete surface from the height
0,7 m.
27
The instrument is proof against static
and variable magnetic fields of strength
up to 400 А/m.
28
The instrument is proof against the
action of radio frequency
electromagnetic fields in the frequency
range 26 – 1000 MHz of strength
up to 10 V/m.
29
The instrument in a transport package is
proof against:
- temperature from minus 50 up to plus
50
0
С;
- humidity up to 100 % at 40
0
С;
- shocks with acceleration 98 m/s
2
,
duration 16 ms;
- vibrations with frequency 10-55 Hz and
bias amplitude 0.35 mm.
30
Weight (with battery) no more than 0.16 kg.
Weight in package no more than 0.3 kg.
31
Dimensions 87х72х35 mm.
32
Reliability parameters:
- average full operating time no less than 20000 h;
- average service life no less than
10 years
- average time of recovery no more than 60 min.
N o t e – For additional information about instrument, please visit www.polimaster.com.
11