EasyManua.ls Logo

Polimaster PM1621MA - 1.3 Specifications

Default Icon
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
1.3 Specifications
1
Operating mode:
- measurement of photon radiation DER;
- measurement of photon radiation DE;
- indication of settings;
- search (for РМ1621М, РМ1621МА);
- indication of partial and critical battery
discharge;
- audio and (additionally for РМ1621М,
РМ1621МА) light and vibro alarm activation
when DE or DER thresholds are exceeded
- indication of the dosimeter’s number ("blind
dosimeter");
- PC data exchange
2
DER measurement range
- for РМ1621, РМ1621М
0.1 Sv/h 100 mSv/h
- for РМ1621А, РМ1621МА
0.1 Sv/h 1.00 Sv/h
DER indication sub-ranges
0,01 9,99 Sv/h;
10,0 99,9 Sv/h;
100 999 Sv/h;
1,00 9,99 mSv/h;
10,0 99,9 mSv/h;
100 200 mSv/h (РМ1621, РМ1621М);
100 999 mSv/h (РМ1621А, РМ1621МА);
1,00 2,00 Sv/h (РМ1621А, РМ1621МА).
- DER analogue scale (seven
segments) in a logarithmic gauge
Number of indicated segments (from left to
right) corresponds to the following DER
threshold values on the LCD display:
- one segment 0.1 Sv/h;
- two segments 1 Sv/h;
- three segments - 10 Sv/h;
- four segments 100 Sv/h;
- five segments 1 mSv/h;
- six segments 10 mSv/h;
- seven segments 100 mSv/h
3
Maximum permissible intrinsic relative
error of DER measurement
(15 + К
1
/ + К
2
) %,
where DER value, mSv/h;
К
1
coefficient 0.0015 (mSv/h);
К
2
coefficient 0.01 (mSv/h)
-1
4
DE measurement range
1.0 Sv 9.99 Sv
5
Maximum permissible intrinsic relative
error of DE measurement
15%
6
DE measurement range
0.01 Sv 9.99 Sv
Indication sub-ranges:
0.01 9.99 Sv;
10.0 99.9 Sv;
100 999 Sv;
1.00 9.99 mSv;
10.0 99.9 mSv;
100 999 mSv;
1.0 Sv 9.99 Sv
7
Discreteness of DE accumulation time
indication
1 h

Related product manuals