RIGOL
Declassification Guide for Digital Oscilloscope
DS1000D Series
DS1000D series digital oscilloscope consists of DS1102D and DS1052D.
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the
size of memory, how it is used, its location, volatility and the clearing procedure.
Instrument memory:
Memory
type and
size
Writable
during
normal
operation
Data
retained
when
powered
off
Purpose/
contents
Data
input
method
Location in
instrument
and remarks
Sanitization
procedure
SDRAM
16MB
Yes No Program Boot Main board in
CPU area
Cycle power
NOR
FLASH
8MB
No Yes Firmware,
Bin,
Internal
saving
Firmware
update,
Internal
saving,
ref/setup
/pf/
record
/wfm
Main board in
CPU area
Save files to
inter
FRAM
1kB
Yes Yes Settings Program Main board in
CPU area
Default
SRAM
512kB
Yes No wfm points Acquire Main board in
acquire area
Cycle power
SRAM
512kB
Yes No LA points Acquire Main board in
acquire area
Cycle power