40
GB
The LED illumination device is in compliance with the
directives of the European Community:
93/42/EEC (Medical Device Directive)
Class I medical product according to Annex IX, Rule 12
(see chap.11)
For the USA: FDA classification Class I
3.6. Information on EMC
(Electromagnetic
Compatibility)
Guidance and manufacturer's declaration – electromagnetic emissions
The EyeMag Light II is intended for use in the electromagnetic environment specified below. The
customer or the user of the EyeMag Light II should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
Radiated Emission
(interference field
strength)
Group 1
Class B
The EyeMag Light II uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Conducted Emission
(interference voltage)
Group 1
Class B
The EyeMag Light II uses RF energy only for its internal
function. Therefore, its RF emissions in charge mode are
very low and are not likely to cause any interference in
nearby electronic equipment.
Harmonic emissions
IEC 61000-3-2
Not applicable The EyeMag Light II is battery powered.
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable The EyeMag Light II is battery powered.