Diagnostics and troubleshooting
9.6Return procedure
Fault description Cause Rectification
Various echoes from an uneven
medium surface, e.g. an extraction
funnel
Check parameter "Material Type"
and adapt, if necessary
Measured value fluctuates
around 10…20%
Reflections from the medium
surface via the vessel wall
(deflection)
Optimize installation position and
transmitter orientation
9.6 Return procedure
Enclose the delivery note, the return goods delivery note and the decontamination
declaration in a clear plastic pouch and attach it firmly to the outside of the
packaging. Any devices/replacement parts which are returned without a
decontamination declaration will be cleaned at your expense before further
processing.
Required forms:
• Delivery note
• Return goods delivery note with the following information: https://
www.siemens.com/processinstrumentation/returngoodsnote
• Product (item description)
• Number of returned devices/replacements parts
• Reason for returning the item(s)
• Decontamination declaration
• https://www.siemens.com/sc/declarationofdecontamination
With this declaration you warrant that the device/replacement part has been carefully
cleaned and is free of residues. The device/replacement part does not pose a hazard
for humans and the environment.
If the returned device/replacement part has come into contact with poisonous,
corrosive, flammable or water-contaminating substances, you must thoroughly
clean and decontaminate the device/replacement part before returning it in order to
40
SITRANS LR100
Operating Instructions, 10/2021, 62705-76-01