Basic
Applications
--
2221
Operators
Л
П
П
O
6.
Set
the
SEC
/
DIV
switch
to
obtain
a
display
of
the
measurement
points
on
the
two
pulses
between
which
the
measurement
is
to
be
made
.
PHASE
DIFFERENCE
BETWEEN
SINUSOIDAL
SIGNALS
1.
Preset
instrument
controls
and
obtain
a
baseline
trace
.
7.
Set
the
trigger
position
as
required
to
obtain
the
entire
pulse
display
.
8.
Press
the
SAVE
/
CONT
switch
to
activate
SAVE
mode
to
save
the
waveform
and
to
present
a
more
stable
display
for
measurement
.
Cursors
will
appear
on
both
the
Channel
1
and
Channel
2
traces
.
2.
Using
probes
or
coaxial
cables
with
equal
time
delay
,
connect
the
reference
signal
to
the
CH
1
OR
X
input
connector
and
the
other
(
phase
-
shifted
)
signal
to
the
CH
2
OR
Y
input
connector
.
3.
Select
a
VERTICAL
MODE
of
BOTH
and
ALT
or
CHOP
,
depending
on
the
input
signal
frequencies
.
9.
Use
the
CURSORS
control
to
move
the
active
cur
sor
to
the
50
%
point
of
the
Channel
1
pulse
leading
edge
.
П
Л
1
7
4.
Set
the
INT
switch
to
the
CH
1
position
and
adjust
the
TRIGGER
LEVEL
control
and
the
TRIGGER
SLOPE
control
for
a
stable
,
triggered
display
.
NOTE
Pulses
with
fast
rise
time
have
only
a
few
sample
points
on
the
leading
edge
,
and
it
may
not
be
possible
to
place
the
dot
at
exactly
the
50
%
level
on
the
leading
edge
.
5.
Use
a
SEC
/
DIV
switch
setting
that
displays
about
two
cycles
of
each
input
signal
.
6.
Set
the
STORE
/
NON
-
STORE
switch
to
the
STORE
position
(
button
in
)
.
10.
Push
the
SELECT
C1
/
C2
button
to
select
the
other
cursor
,
and
use
the
CURSORS
control
to
position
the
active
cursor
at
the
50
%
level
of
leading
edge
of
the
Chan
nel
2
pulse
(
see
Figure
6-19
)
.
П
1
1
7.
Check
that
the
TRIGGER
LEVEL
control
is
adjusted
for
a
stable
,
triggered
acquisition
.
11.
Read
the
time
difference
between
the
pulses
from
the
crt
readout
.
NOTE
Use
the
NORM
Trigger
Mode
for
low
-
repetition
-
rate
signals
(
below
approximately
20
Hz
)
.
This
ensures
that
the
storage
window
and
trigger
signal
are
syn
chronized
when
the
trace
is
triggered
.
AV15.80V
AV2
=
2.20V
AT
=
0.149ms
100
90
8.
Set
both
VOLTS
/
DIV
switches
and
adjust
the
VOLTS
/
DIV
Variable
controls
to
obtain
a
5
-
division
vertical
display
of
each
input
signal
.
CHANNEL
1
NOTE
CURSORS
10
CHANNEL
2
П
П
П
Use
the
Vertical
POSITION
controls
in
conjunction
with
the
VOLTS
/
DIV
Variable
controls
to
vertically
center
the
5
-
division
display
between
the
%
and
100
%
dotted
reference
graticule
lines
.
sv
av
Olims
4998-21
9.
Set
the
trigger
position
and
TRIGGER
SLOPE
switch
as
necessary
to
place
the
measurement
points
within
the
graticule
area
(
see
Figure
6-20A
)
.
Figure
8-19
.
Time
difference
between
two
time
-
related
pulses
.
П
6-19