Performance verification
Frequency/pe
riod test
This test verifies the frequency accuracy of the arbitrary function generator. All
output frequencies are derived from a single generated frequency. Only one
frequency po
int of channel 1 is required to be checked.
1. Connect the arbitrary function generator to the frequency counter as shownin
the followi
ng figure.
Figure 3: Frequency/period test
2. Push the Default front-panel button and the OK menu button to recall the
arbitrary function generator default setup.
3. Set up the arbitrary function generator using the following steps:
Select menu Setting Operation
Func
tion
Sine Sine (front)
Freq
uency
1.00
0000 MHz
Frequency/Period (front)
Amp
litude
1.0
0 Vpp
Amplitude/High (front)
Channel 1 Output On On (front)
4. Check that reading of the Frequency Counter is between 0.999998 MHz and
1.000002 MHz.
5. Set up the arbitrary function generator using the following steps:
Select menu Setting Operation
Function Pulse
Pulse (front)
6. Check that reading of the frequency counter is between 0.999998 MHz and
1.000002 MHz.
AFG3000C and AFG3000 Series Specifications and Performance Verification 51
NOTE. Frequency counter values must include stability and aging.

Table of Contents

Other manuals for Tektronix AFG3000C Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3000C Series and is the answer not in the manual?

Tektronix AFG3000C Series Specifications

General IconGeneral
BrandTektronix
ModelAFG3000C Series
CategoryInverter
LanguageEnglish

Related product manuals