Performance verification
Select menu Setting Operation
Offset
0.0 V
TopMenu > Offset/Low (front)
Channel 1 Output On On (front)
3. Set up the Oscilloscope so the square waveform of 5 division amplitude is
displayed.
4. Verify that the rise/fall time of the square waveform on the oscilloscope at
each amplitude is equal to or less than the limit specified in the following
tables. Use 10-90% reference level for rise/fall time measurement.
5. (Two channel models only) Repeat steps 2 through 4 for the channel 2 output.
AFG3011 / 3011C Oscilloscope Measurement
Functio
n
Frequen
cy
Offset
Amplitu
de
Vertica
l
Horizon
tal
Limit
Square
2.00 MHz 0.0 V 1.0 Vpp
200 mV/div 20 ns/div or
25 ns/div
≤ 50 ns
Square
2.00 MHz 0.0 V 10.0 Vpp
200 mV /div
with x1
0
attenuator
20 ns/div or
25 ns/d
iv
≤ 50 ns
AFG302xB Oscilloscope Measurement
Funct
ion
Frequ
ency
Offset
Ampli
tude
Verti
cal
Horiz
ontal
Limit
Square
10.0
0MHz
0.0 V 1.0 V
pp
200 mV/div 5 ns/div
≤ 18 n
s
Square
10.0
0MHz
0.0 V 10.0
Vpp
200 mV/div
with x10
att
enuator
5ns/div
≤ 18 n
s
AFG
302xC
Osc
illoscope
Mea
surement
Function Frequency
Of
fset
Amplitude Vertical Horizontal Limit
Sq
uare
10.00 MHz 0.0 V 1.0 Vpp
20
0mV/div
5n
s/div
≤ 9ns
S
quare
10.00 MHz 0.0 V 10.0 Vpp
2
00 mV /div
with x10
attenuator
5
ns/div
≤ 9ns
AFG305xC Oscilloscope Measurement
Function Frequency
Offset
Amplitude Vertical Horizontal Limit
Square
10.00 MHz 0.0 V 1.0 Vpp
200 mV/div 5 ns/div
≤ 7ns
Square
10.00 MHz 0.0 V 10.0 Vpp
200 mV /div
with x10
attenuator
5ns/div
≤ 7ns
68 AFG3000C and AFG3000 Series Specifications and Performance Verification