Performance verification
Select menu Setting Operation
Function
DC More... (front) > More Waveform
Menu > DC
Offset
+ 5.000 V + 2.500 V
(AFG325x)
Offset/Low (front)
Channel 1 Output On On (front)
5. Verify that each offset measurement is within the range specified in the
following
tables.
6. (Two channel models only) Repeat steps 3 through 4 for the channel 2 output.
AFG3011 / 3011C
Function
Offset
Measurement Range
DC
+ 5.000 Vdc Vdc
(5.000 × CF ± 0.110) Vdc
DC
0.000 Vdc Vdc ± 0.010 Vdc
DC
- 5.000 Vdc Vdc
(- 5.000 × C F ± 0.110) Vdc
AFG302xB / 302xC / 305xC / 310x / 310xC / 315xC M easurem ent Range
Function
Offset
DC
+ 5.000 Vdc Vdc
(5.000 × CF ± 0.055) Vdc
DC
0.000 Vdc Vdc ± 0.005 Vdc
DC
- 5.000 Vdc Vdc
(- 5.000 × CF ± 0.055) Vdc
AFG325x / 325xC
Function
Offset
Measurement Range
DC
+ 2.500 Vdc Vdc
(2.500 × CF ± 0.030) Vdc
DC
0.000 Vdc Vdc ± 0.005 Vdc
DC
- 2.500 Vdc Vdc
(- 2.500 × CF ± 0.030) Vdc
AFG3000C and AFG3000 Series Specifications and Performance Verification 55
4. Set
up
the arbitrary function generator using the following steps:

Table of Contents

Other manuals for Tektronix AFG3011

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix AFG3011 and is the answer not in the manual?

Tektronix AFG3011 Specifications

General IconGeneral
BrandTektronix
ModelAFG3011
CategoryInverter
LanguageEnglish

Related product manuals