Operating basics
1. In this example
,a
5 M Hz continuous
sine waveform is used.
Conrm that bo
th phases
are set to 0 degrees.
2. Change the C
H1
frequency to 10 MHz,
and then back to 5 MHz.
In this sta
te, the CH2
phase does not return to
its initial condition.
3. To align t
he phase of two
channel signals, push the
Align Phase bezel button.
4. When you
push the
Align Phase button,
the instrument will stop
the sig
nal generation,
adjust the phases of
both channels, and
automa
tically restart the
signal generation.