EasyManuals Logo
Home>Tektronix>Measuring Instruments>CSA7404

Tektronix CSA7404 User Manual

Tektronix CSA7404
480 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #261 background imageLoading...
Page #261 background image
Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3- 183
Overview Related control elements and resourcesTo deskew channels (Cont.)
Compensate
probe timing
(deskew)
(Cont.)
12. Touch Deskew Time and use the mul tipurpose knobs or
keypad to adjust the deskew time for that channel so
that its signal aligns with the trigger position.
13. Repeat steps 11 and 12 for each additional channel that
you want to deskew.
14. Remove the connections.
For further
assistance
15. Touch the Help button to access the online assistance.
See page 3--283 to learn about using online help.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404 and is the answer not in the manual?

Tektronix CSA7404 Specifications

General IconGeneral
BrandTektronix
ModelCSA7404
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals