Built-In Security Features
When to Use Tek
Secure
You can use the TekSecure function to erase setup and reference waveform data
stored in internal ash memory.
The DPO2000 and MSO2000 series models have a USB host port on the front
panel. Any USB ash d evices can be removed, and stored or destroyed.
What TekSecure Does
The TekSecure function does the following:
Replaces all waveforms in all reference memories with null sample values
Replaces the current front-panel setup and all stored setups with the default
setup values
Calculates the checksums of all reference waveform memory an d setup
memory locations to verify successful completion of waveform and setup
erasure
Displays a dialog indicating whether the secure erase was successful or
unsuccessful
NOTE
. TekSecure does not erase or change factory calibration constants or
Ethernet settings.
To u
se TekSecure:
1. Push the front-panel Utility button.
2. Push the Utility Page lower-bezel button a nd use the Multipurpose a knob
to select Cong.
3. Push the TekSecure Er ase Memory lower-bezel button.
4. Push the OK Erase Setup & Ref Memory side-bezel button. Wait for the
“TekSecure operation complete” dialog box to d isplay.
5. Push the Menu Off front-panel button to close the dialog box.
DPO2000 and MSO2000 Series Declassication and Security Instructions 5

Other manuals for Tektronix DPO2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2000 Series and is the answer not in the manual?

Tektronix DPO2000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals