Clear and Sanitize Procedures
Memory Devices
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options.
Table 1: Volatile Memory Devices for the DPO2000 and MSO2000 Series
Type and minimum size Function User modiable
Data input
method Location To clear or sanitize
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the analog channel
waveform acquisition
data during operation
Only by the
processor system
N/A
Acquisition board
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the digital channel
waveform acquisition
data during operation
Only by the
processor system
N/A
Acquisition board
(MSO2000 Series only)
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 64MX16,
1G DDR2, 1.8V, 333MHZ;
MT47H64M16BT-3, FBGA84
DDR memory: contains
the instrument code and
data during operation
Only by the
processor system
N/A
Acquisition board
Remove power from the
instrument for at least
20 seconds
CMOS, SDRAM, 32MX16,
512MB DDR2, 1.8V, 333MHZ;
MT47H32M16CC-3, FBGA84
DDR memory: contains
the instrument code and
data during operation
Only by the
processor system
N/A
Acquisition board
(4-channel products
only)
Remove power from the
instrument for at least
20 seconds
Table 2: Nonvolatile Memory Devices for the DPO2000 and MSO2000 Series
Type and minimum size Function User modiable
Data input
method Location To clear or sanitize
CMOS,FLASH;16MX8OR
8M X 16, 128MBIT, 75NS, 3V;
PC28F128J3D-75, BGA64
FLASH: contains the
instrument rmware,
calibration constants,
reference waveforms,
and instrument setups
N/A N/A
Acquisition board To clear: Not applicable
for calibration constants.
Use the TekSecure
feature to erase
reference waveforms
and instrument setups.
Use the procedure to
disable the LAN Ethernet
and to clear information,
such as IP addresses.
(See page 6, Disable the
LAN Port and Clear LAN
Ethernet Settings.)
To sanitize: Remove
the acquisition board;
see the product service
manual for removal
instructions. Store the
removed board in a
secure area or destroy
it. Return the product
to Tektronix and a
new acquisition board
will be installed and
the instrument will be
repaired and adjusted as
necessary.
DPO2000 and MSO2000 Series Declassication and Security Instructions 1

Other manuals for Tektronix DPO2012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012 and is the answer not in the manual?

Tektronix DPO2012 Specifications

General IconGeneral
BrandTektronix
ModelDPO2012
CategoryTest Equipment
LanguageEnglish

Related product manuals